N388PH

EUROCOPTER DEUTSCHLAND GMBH EC 135 P2+ · 2008 · Valid Registration

Registered owner: PHI HEALTH LLC, AZ (owner ≠ operator)

Operators of this aircraft

Who operated this tail, and how firmly we know it.

OperatorRolePeriodBasis
PHI HEALTH,LLC (VE7A)Certificate holderMatched by certificate designator

Accident & incident history

No NTSB accidents or incidents on file for this aircraft.

Maintenance disclosures

B · 2018-01-07 Matched by certificate designator
ATA 2422

PILOT AND MEDICAL PERSONNEL ALL SMELLED SOMETHING BURNING AND THE ODOR BECAME MORE INTENSE IN THE AFT CABIN. ONE OF THE MEDICAL PERSONNEL STARTED TO SEE SMOKE IN THE AFT CABIN AREA AND NOTICED A FLICKERING OF THE LIGHTING ON THE MEDICAL EQUIPMENT. SWITCHED THE EMS 110VAC INVERTER OFF AT THAT TIME. TURNED OFF THE AIR CONDITIONER AND MEDICAL OXYGEN. OPENED UP THE SLIDE WINDOW AND VENTS. THE SMOKE AND ODOR STARTED TO DISSIPATE ON DESCENT. AT LANDING, INITIATED AN EMERGENCY SHUTDOWN PROCEDURES. STATIC INVERTER FAILED.

Source: SDR VE7A2018010700001 · FAA SDRS